Operating
+ 10 °C to + 40 °C, < 90 % relative humidity
conditions
(non-condensing), 800 –1050 hPa ambient
pressure
Storage and
-20 °C to + 55 °C, < 90% relative humidity
transport
conditions
Power supply 4 x 1.5 V
Battery life
For approx. 300 measurements, depending on
blood pressure and inflation pressure levels
Product life
Information on the life cycle of the product can
cycle to be
be found at beurer.com
expected
Classification Internal power supply, IP 21 no AP or APG,
continuous operation
Blood pressure: Application part, type BF
The serial number is located on the device or in the battery com-
partment.
We reserve the right to make technical changes to improve and
develop the product.
• This device conforms with the European standard
EN 60601-1-2 (Group 1, Class B, in accordance with CISPR-11,
IEC 61000-3-2, IEC 61000-3-3, IEC 61000-4-2, IEC 61000-4-3,
IEC 61000-4-4, IEC 61000-4-5, IEC 61000-4-6, IEC 61000-4-7,
IEC 61000-4-8, IEC 61000-4-11) and is subject to particular
precautions with regard to electromagnetic compatibility.
Please note that portable and mobile HF communication sys-
tems may interfere with this device.
• The accuracy of this blood pressure monitor has been carefully
checked and developed with regard to a long useful life. If the
device is used for commercial medical purposes, the applica-
LR03 AAA batteries
ble national regulations determine whether it must be tested
for accuracy by appropriate means.
Mains adapter
Model no.
LXCP12X-050100BG
Input
100 – 240 V, 50 – 60 Hz, 0.5 A max
Output
5 V DC, 1 A, in conjunction with Beurer blood
pressure monitors only
Manufacturer
Shenzhen Longxc Power Supply Co., Ltd
Protection
This device is double-insulated and has a pri-
mary-side cutout switch which disconnects the
device from the mains in case of malfunction.
Make sure that you have removed the batteries
from the battery compartment before you use
the mains adapter.
Polarity
Insulated/protection class 2
Housing and
The housing of the mains adapter protects us-
protective
ers from touching parts that are or could be live
covers
(for example with their fingers, or with a needle
or checking hook).
The user must not touch the patient and the
output connector of the AC/DC mains adapter
at the same time.
12. GUARANTEE/SERVICE
Further information on the guarantee and guarantee conditions
can be found in the guarantee leaflet supplied.
27