6. SYSTEM ERRORS
Like the stretcher, the rail has possible errors related to various causes.
Below are the common errors of the rail system:
DESCRIPTION
CODE
OF THE ERROR
1
The stretcher does not
receive voltage from the
ambulance.
2
Failure in charging process.
REV 2
10-24
TROUBLESHOOTING
POSSIBLE CAUSE OF THE
ERROR
The stretcher-rail plug is not
connecting correctly
Ambulance battery is discharged
or damaged.
Power cables incorrectly
connected either on the
ambulance, rail or stretcher.
One of the copper contacts of the
stretcher plug has not come out
of its housing completely in the
plastic part.
Contact technical support
Instruction manual
SUPERBRAVO 1 / 2
ERROR
SOLUTION
Check that the two contacts of
the stretcher plug and the rail are
touching each other and that there is
no foreign body between them.
Charge the battery or replace it with
a new one, respectively.
Check connection of the cables.
Recommended order of testing:
ambulance, rail and stretcher.
Check that the plug contacts move
smoothly in and out of the socket.
Contact technical support.
41