limescale formation on the mesh. The parts should not come into contact with the hot base of the
pan.
• Then dry all parts thoroughly. For more information, see the "Drying" section.
• The medication container
• We guarantee 52 disinfection cycles with boiling water per medication container.
NOTICE
The masks must not be placed in hot water!
• We recommend disinfecting the masks with 70–75% ethyl alcohol.
Drying
• Dry the parts carefully using a soft cloth.
• Gently shake the medication container
holes in the mesh.
• Place the individual parts on a clean, dry and absorbent surface and leave them to dry completely
(for at least 4 hours).
• Do not use a microwave, dish dryer or hair dryer for drying or disinfection.
NOTICE
Please ensure that the parts are completely dried after cleaning, otherwise there is a high risk of bacterial
growth.
When completely dry, reassemble the parts and place in a dry, sealed container.
Durability of materials
• Like all plastic parts, the medication container
degree of wear and tear if used frequently and prepared hygienically. Over time, this may lead to a
change in the aerosol, which may reduce the e ectiveness of the treatment. We therefore recom-
mend that you replace the medication container
must not be placed in the microwave.
2
a few times so that any water is removed from the tiny
2
54
and the atomiser set are subject to a certain
2
and atomiser set after one year at the most.
2