SYNCHRONIZATION
If the automatic mode does not work properly, follow
this calibration procedure:
1. Connect the power cable of the micromotor to one of
the rear power sockets, which are marked with
numbers 1, 2, 3.
2. Turn on the micromotor and set it to spin at
10 000 rpm but don't start it yet.
3. From the home screen, press and hold the key
CUSTOMER .CONFIG
3.0
User
4.0
Devices.
4. Pulse la tecla
para seleccionar el menú 2.0
Dispositivos y vuelva a pulsar brevemente la tecla
.
4.0 Devices
2.1
Device
1.
2.2 Device 2
2.3 Device 3
Selecct device 1, 2 or 3 with the key
key
.
5. HBIL: Enable device.
 YES: Automatic detection enabled.
 NO: Automatic detection disabled.
It is recommended to disable sockets that are not
going to be used, to eliminate possible interference.
2.1 Device 1
HBIL = .YES.
GAIN = LO
CNmot = NO VNmot = 3000
Vnmot = 10 Vmot = 0
6. GAIN: Configurable sensitivity.
 LO: Low sensitivity. It is recommended when
connecting a utensil for the first time.
 HI: High sensitivity. Use only if satisfactory
results are not achieved with the LO value.
When changing the sensitivity, the device will need
10 seconds to reconfigure. The GAIN ADJ. message
will flash.
2.1 Device 1
HBIL
= YES
GAIN = LO
CNmot = NO VNmot = 3000
Vnmot = 10
Vmot = 0
GAIN
ADJ.
7. To calibrate a micromotor,
8. The display will show the message "ACTIVATE".
.
9. After a few seconds, the display will show the
. Press the
10. The aspiration will start for a few seconds and one of
11. If the device detection after calibration is not
12. Select another micromotor to calibrate or abandon the
- access the parameter CNmot with the key
- select the value YES with the key
- and press the key
Immediately run the micromotor at 10 000 rpm.
2.1 Device 1
HBIL
= YES
GAIN = LO
CNmot = NO VNmot = 3000
Vnmot = 10
Vmot = 0
ACTIVATE.
message "RELEASE" while emitting an acoustic
signal. Now stop the micromotor. Immediatly stop
the micromotor.
Device
2.1
1
HBIL
= YES
GAIN = LO
CNmot = NO VNmot = 3000
Vnmot = 10
Vmot = 0
RELEASE.
the following messages will appear:
"CAL OK": Calibration is successful.
"ERROR!": The calibrated utensil will probably not
be correctly detected automatically.
Electrical noise can be excessive due to
a bad ground or because the level to be
calibrated is too low.
Device
2.1
1
HBIL
= YES
GAIN = LO
CNmot = NO VNmot = 90
Vnmot = 10
Vmot = 15
CAL
OK.
satisfactory, it is recommended:
 Repeat the process several times.
 Increase the speed of the device in steps of
2000 rpm and repeat the calibration.
 Set the sensitivity GAIN = HI, set the device to
10 000 rmp and repeat the calibration.
process by pressing the key
.
.
,