Test Standard Test Item
IEC 61000-3-2 Harmonic Current Emission
IEC 61000-3-3 Voltage Fluctuations and Flicker Not applicable (Note 1) –
Note 1: The power supply of this device is from battery (DC voltage)
A1.3 Guidance and manufacturer's declaration – Electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure
it is used in such an environment.
Test Standard
Test Item
IEC 61000-4-2
Electrostatic discharge
IEC 61000-4-3
Radiated RF EM fields
IEC 61000-4-3
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-4
Electrical fast transients /
bursts
IEC 61000-4-5
Surges line-to-line
IEC 61000-4-6
Conducted disturbances
induced by RF fields
IEC 61000-4-8
Rated power frequency
magnetic fields
Test Level
Not applicable (Note 1) –
Test Level
Electromagnetic environment – Guidance
Contact:
Floors should be wood, concrete or ceramic tile.
± 8 kV;
If floors are covered with synthetic material, the
Air: ±2 kV, ±4 kV,
relative humidity should be at least 30%.
±8 kV, and ±15 kV
Portable and mobile RF communications
equipment should be used no closer than 30 cm
(12 inch) to any part of this device.
10 V/m;
Portable and mobile RF communications
80 MHz to 2.7 GHz;
equipment should be used no closer than 30 cm
80% AM at 1 kHz
(12 inch) to any part of this device.
Not applicable
(Note 1)
Not applicable
(Note 1)
Not applicable
(Note 1)
Not applicable
(Note 1)
30 A/m
Power frequency magnetic fields should be at
50 Hz and 60 Hz
levels characteristic of a typical location in a
typical commercial or hospital environment.
71
Electromagnetic environment – Guidance

Publicité