• Do not expose the device to direct sunlight or high
temperatures.
• Protect the device from dust, dirt and humidity.
• Never immerse the device in water or other liquids.
• For hygienic reasons, use the gel pads only on one person.
• The device is suitable for self-treatment.
• Switch off the device immediately if it malfunctions or if you
feel unwell or experience pain.
• Switch off the unit or the respective channel first before
removing or relocating electrodes to prevent unintentional
stimulation.
• Do not modify electrodes (e.g. by cutting them), as
this increases the current density, which is potentially
hazardous (max. recommended output value for the
electrodes is 9 mA/cm², an effective current density
beyond 2 mA/cm² requires increased awareness).
• Never directly contact the electrode area or apply the
electrode to any part of the body. This product must be
used in conjunction with a set of self-adhesive electrode
gel pad. Please follow the steps in "Initial Use" to install
the self-adhesive electrode gel pads to the electrode areas
before use.
• Do not use when asleep, when driving a vehicle or when
operating machinery.
• Do not use whilst undertaking any activity where an
unexpected reaction (e.g. strong muscle contractions even
at low intensity) could be dangerous.
• Ensure that no metallic objects come into contact with the
electrodes during stimulation. Remove all jewelry or body
piercings before using the device. Failure to do so could
result in spot burns.
• Make sure not to confuse the electrode cables including
contacts with your headphones or other devices, and do
not connect the electrodes to other devices.
• Do not use this device at the same time as other devices
that emit electrical pulses.
• Do not use in the vicinity of highly flammable substances,
gases or explosives.
• The actual temperature may vary depending on the
condition of your skin, your age, the location of the pain, etc.
• If the heat function feels too hot, stop treatment immediately.
You can continue the TENS or EMS treatment without the
heat function.
• During the initial few minutes, use the device while sitting
or lying down to minimise the risk of injury resulting from
isolated cases of vagal responses (feeling of faintness). If you
feel faint, immediately switch off the device, lie down and
support your legs in an elevated position (approx. 5–10 min.).
• Treatment of the skin with moisturising lotions or ointments
beforehand is not recommended as this considerably
increases the gel pad wear and may cause unpleasant
current peaks.
• This device is not intended for use by children or people
with restricted physical, sensory (e.g. reduced sensitivity to
pain) or mental skills or a lack of experience and/or lack of
knowledge, unless they are supervised by a person who is
responsible for their safety or who are instructed by such
a person in how to use the device.
• If the adhesive capability of the gel pads decreases, please
replace them immediately. Do not use the device until you
have the new gel pads. Otherwise the unequal adhesion of
the gel pads may lead to skin injuries. Replace the gel pads
with new ones at the latest after having used them 20 times.
5