EMS 34
In this program, you can set the impulse frequency between 1
and 150 Hz and the impulse width between 80 and 450 μs. You
can also set the working time and pause time for this program
between 1 and 30 seconds each.
1. Place the electrodes on the desired area for treatment (for
placement suggestions see the "Electrode Positioning"
section) and connect them to the device.
2. Select the EMS 34 program as described in the "Starting
Application" section, steps 3-5.
˄/˅
3. Use the
setting buttons to select the working time ("on
time") and use the ENTER Button to confirm.
˄/˅
4. Use the
setting buttons to select the pause time ("off
time") and use the ENTER Button to confirm.
˄/˅
5. Use the
setting buttons to select an impulse frequency
and use the ENTER Button to confirm.
˄/˅
6. Use the
setting buttons to select an impulse width and
use the ENTER Button to confirm.
˄/˅
7. Use the
setting buttons to select the treatment time and
use the ENTER Button to confirm.
˄/˅
8. Use the left and right
to select an impulse intensity.
EMS 35
Various impulse sequences run in this burst program. Burst
programs are suitable for all areas of application to be
treated with changing signal patterns (to minimize the level
of accustoming to the treatment). In this program, you can
set the impulse frequency between 1 and 150 Hz and the
impulse width between 80 and 450 μs. You can also set the
working time and pause time for this program between 1
and 30 seconds each.
setting buttons for CH1 and CH2
1. Place the electrodes on the desired area for treatment (for
placement suggestions see the "Electrode Positioning"
section) and connect them to the device.
2. Select the EMS 35 program as described in the "Starting
Application" section, steps 3-5.
˄/˅
3. Use the
setting buttons to select the working time ("on
time") and use the ENTER button to confirm.
˄/˅
4. Use the
setting buttons to select the pause time ("off
time") and use the ENTER button to confirm.
˄/˅
5. Use the
setting buttons to select an impulse frequency
and use the ENTER button to confirm.
˄/˅
6. Use the
setting buttons to select an impulse width and
use the ENTER button to confirm.
˄/˅
7. Use the
setting buttons to select the treatment time and
use the ENTER button to confirm.
8. Use the left and right
to select an impulse intensity.
11. Doctor's Function
The Doctor's Function is a special setting to allow you to
access your personal program even more easily and directly.
Your individual program settings are instantly recalled and
activated when the device is switched on. You may wish to
adjust this individual program following advice from your
doctor.
Setting the Doctor's Function
1. Select your program and the corresponding settings as
described in the "Starting Application" section.
2. At the start of the stimulation treatment, the impulse intensity
of CH1 and CH2 is set to 00 by default. No impulses are
sent to the electrodes yet. Before setting the desired impulse
16
˄/˅
setting buttons for CH1 and CH2