10.5 DC POWER CORD FUSE REPLACEMENT (RP-125)
The cigarette lighter DC power cord contains a fuse. If the DC power cord is being used with a known good power
source and the device is not receiving power, the fuse may need to be replaced.
WARNING
CHOKING HAZARD: small parts exposed when changing the fuse, keep away from small children and pets.
• CRITICAL FUSE SIZING: incorrect fuse replacement size may result in fire or inadequate equipment protec-
tion. Replace only with same type and rating of fuse.
• ELECTRICAL SHOCK: complete disconnect the cable before attempting to change the fuse.
• Do not hang any type of accessory or accessory bracket from plug.
To replace the fuse:
Step
Instruction
1.1 Remove the tip by unscrewing the
1
retainer. Use a tool if necessary.
2
2.1 Remove the retainer, tip, and fuse.
3.1 The spring should remain inside the
3
cigarette lighter adaptor housing.
3.2 If the spring is removed, replace
the spring first before inserting the
replacement fuse.
4.1 Install a replacement fuse.
4
4.2 Reassemble the tip.
4.3 Ensure the retainer ring is properly
seated and tightened.