The maximum permitted temperature inside the reac-
tor vessel is 230 °C. Do not exceed this temperature.
Before you fill the reactor vessel, ensure that the
reagents used do not corrode the seal!
Accessories
Only use IKA approved accessories!
Use only original IKA spare parts!
Protect the instrument and accessories from bumping
and impacting.
Please observe the permitted speed for the stirring
element. Never set higher speed.
Assembly
Ensure that the stirring tool is
WARNING
securely fixed.
Broken glass! Fit the hose cou-
CAUTION
pling carefully.
• Fix all hoses in the laboratory extractor hood.
• Ensure that the glass assembly is tension-free! Danger
of cracking as a result of:
- stress due to incorrect assembly,
- external mechanical hazards,
- local temperature peaks.
16
Stand
Undo the screws on the stand
CAUTION
carefully.
balance the stand can move
upwards uncontrollably.
Protection of the device
The voltage stated on the type plate must correspond
to the mains voltage.
Socket must be earthed (protective ground contact).
Covering or parts that are capable of being removed from
the unit without accessory equipment have to be reat-
tached to the unit for safe operation in order to prevent,
for example, the ingress of fluids, foreign matter, etc..
The device may only be opened by experts.