The anchor stirrer must only be
NOTICE
operated at a speed of max.
300 rpm.
You can use these in conjunction with labworldsoft
ratory software to create a high-performance system for
measuring, controlling, regulating and fully documenting
all laboratory processes.
Sample applications:
Manufacture of cremes, lotions, emulsions and liposo-
me preparations in the pharmaceutical and cosmetic
sector (lipstick, sun cream, eyeliner).
Unpacking
- Unpack the device carefully
- Any damage should be notified immediately to the ship-
ping agent (post office, railway network or transport
company)
The laboratory reactor system is supplied as a fully functi-
onal unit in various configurations depending on customer
requirements. Accessories, such as ULTRA-TURRAX, tempe-
rature gauges and reactor vessels are supplied separately.
LR-2.ST / LR-5.ST stand system
The stand system holds the
NOTICE
drive unit complete with the
reactor cover and the stirring
instrument, the reactor vessel
and the safety disconnector.
An adjustment screw on the rear, right-hand side foot of
the stand can be used to level the stand and compensate
for any unevenness in the mounting surface.
Undo the lock nut. Turn the knurled screw clockwise to
slightly raise the foot of the device, turn it anticlockwise
to lower it. When the device is level, tighten the lock nut.
Intended use
®
labo-
Unpacking
Scope of delivery
- LR-2.ST or LR-5.ST stand system complete with reactor cover
- Overhead stirrer
- LR-2.SI safety disconnector
Assembling
If you are working with the LR 2.1 single-walled reactor
vessel and the HBR 4 digital heating bath, you must remo-
ve the baseplate including clamping bracket and one side
support from the stand.
Undo the top two M8 cheese-head screws using the DIN 911-6
cranked wrench key provided. Remove the baseplate, undo the
bottom cheese-head screws and remove the side support.
The remaining side support is used to fit the safety disconnec-
tor. Place the plastic cover onto the aluminium profile.
M8 cheese-head
screw with basepla-
te and side support
Adjuster
screw for
levelling
Mixing of solids such as calcium carbonate, talc and
titanium oxide with liquid polymers.
Mixing of additives and solid polymer compounds with
mineral oils.
Dispersion and mixing of solids and fibres in liquids
and polymers.
Manufacture of milk products and dietary products.
Extraction of enzymes from biomass.
Undo the knur-
led screw to
adjust the height
17