Example: when the data is 25 November 2013, the name of the folder will be: 20131125.
The data will be written as a text file with a name that consists of the time in 24h format (hr min sec)
with the prefix S.
Example: when settings are exported at 15:12:25 (3:12:25 pm) the name of the file will be:
S151225.txt
Note
Pressing EXIT during the exporting aborts the process
3. Factory Reset
When the factory reset is performed, the instrument will revert back to the original settings when the
instrument left the factory. All data will be lost and the PIN is set-back to the original PIN "000000".
5.8 Instrument self-test
The instrument self-test requires user interaction.
1
In the Setup menu, select "6. Instrument Self-test".
Selecting the menu item starts the self-test routine.
2
Press the function keys on the keypad one by one in any order.
The self-test result is displayed after a few seconds.
The meter returns to the system settings menu automatically.
Note
●
The user needs to finish pressing all seven keys within two minutes, otherwise "Self-test failed!"
appears and the procedure has to be repeated.
●
If error messages repeatedly appear, contact METTLER TOLEDO Service.
26
Setup