2. Time and date
When starting the meter for the first time, the display for entering time and date appears automatically.
In the system settings, two time and four date display formats are available:
●
Time
24-hour format (for example, 06:56 and 18:56)
12-hour format (for example, 06:56 AM and 06:56 PM)
●
Date:
28-11-2013 (day-month-year)
11-28-2013 (month-day-year)
28-Nov-2013 (day-month-year)
28/11/2013 (day-month-year)
3. Access control
PIN settings are available for:
1. System settings
2. Deleting data
3. Instrument login
1
Switch PIN protection for the required access control ON. The window for entering an alphanumer
ic PIN appears.
2
Enter an alphanumeric PIN (max. 6 characters).
The input window for PIN verification appears.
3
Confirm PIN.
A maximum of 6 characters can be entered as PIN. In the factory default settings, the PIN for system
settings and deleting data is set to 000000 and is activated, no instrument login password is set.
4. Beep
An acoustic signal can be switched on in the following three cases:
1. Key is pressed
2. Alarm/warning message appears
3. Measurement is stable and has endpointed (stability signal appears)
5. Expert/Routine modes
The meter has two working modes:
●
Expert mode: the factory default setting enables all functions of the meter.
●
Routine mode: some of the menu settings are blocked.
The concept of the two working modes is a GLP feature that ensures that important settings and stored
data cannot be deleted cannot be unintentionally changed under routine working conditions.
The meter only allows the following functions in the routine mode:
●
Calibrating and measuring
●
Editing user, sample and sensor IDs
●
Editing the MTC temperature
●
Editing data transfer settings
●
Editing system-settings (PIN-protected)
●
Running the instrument self-test
24
Setup