2. Interface
Select to transfer the data in the memory to a Printer, LabX Direct or both. The meter adjusts the baud
rate to the following settings in case no automatic baud rate synchronization occurs (only possible
with USB and the printers RS-P25, RS-P26 and RS-P28):
1. Printer
Baud rate: 1200
Data bits: 8
Parity: none
Stop bits: 1
Handshake: none
2. LabX direct
The settings between the instrument and PC are adjusted automatically because USB connection
is plug-and-play
3. Printer + LabX direct
The settings listed above for 1. and 2. are used.
3. Printout Format
Three different printout formats are available: GLP, Normal and Short. The printouts can be printed in
six different languages, depending on which language is currently selected in Setup (English, German,
French, Italian, Spanish and Portuguese). For all other languages the printouts will be in English.
* If LabX direct is selected, the printout format is always GLP and English. LabX direct PC software
translates the received data into the selected PC language as defined in the regional and language
options.
Examples:
pH printout GLP
<pH>
GLP
24-Jul-13
10:56 AM
BEER
6.997 pH
0.5 mV
25.0 °C ATC
Auto EP Strict
INLAB413
122222222
Last cal.: 09-Jun-2013
10:56 AM
Ivy
Signature:____________
Outside limits
5.6 System settings
The system settings menu is protected by a PIN. Upon delivery, the PIN is set to 000000 and is acti
vated. Please change the PIN to prevent unauthorized access.
1. Language
The following languages are available for the system: English, German, French, Spanish, Italian, Por
tuguese, Chinese, Japanese, Korean and Russian.
pH printout normal
<pH>
Normal
24-Jul-13
10:56 AM
BEER
6.997 pH
0.5 mV
25.0 °C ATC
Auto EP Strict
INLAB413
pH printout short
<pH>
0.5 mV
25.0 °C ATC
Auto EP Strict
INLAB413
Setup
23