HF current does not flow when the foot switch is actuated. The system only reacts with a
tone signal and the display changes colour.
Please check:
➔
That all terminals have been correctly connected
➔
The settings of the foot switch
➔
Whether the neutral electrode has been recognized
◼
The neutral electrode is not or not correctly detected
Check all terminals and touch the field with the symbol of the neutral electrode for detection of the
NE. If the neutral electrode is still not recognized, exchange it where appropriate.
◼
It is not possible to switch the machine on or off.
Check whether the power cable has been correctly connected and that the socket is intact. If the
error persists, inform the specialist personnel.
◼
The system works perfectly, but the settings cannot be saved.
Inform the specialist personnel.