HBS x-touch 300 argon BiSeal
Areas for attachment of the neutral electrode
Attach the neutral electrode to clean, dry skin.
The electrode must be attached close to the operation area, but with a minimum direct distance from
the operating field of 20 cm.
Attach over tissue with a good supply of blood without dips or curves in the skin, for instance on the
upper arm or thigh. (cf. picture below)
Do not attach the neutral electrode to projecting surfaces.
Do not attach above bones, scars, cuts or scratches.
The electrode must not be attached to surfaces with a high amount of hair growth. If necessary,
shave the relevant area before attaching.
Do not attach above particularly adipose areas such as the abdomen or buttocks.
The neutral electrode must not be attached above implants.
Electrodes of monitoring systems must be placed at a distance of at least 20 cm from the operation
area and from the position of the attached neutral electrode
Correct attachment of the neutral electrode in adult patients
Correct attachment of the neutral electrode in children
Attachment in patients with pacemakers
Note! Consult the treating cardiologist!
®
RIGHT
WRONG!