Commissioning the unit
8.1 Before starting work
Before starting surgery, the HBS x-touch 300 argon BiSeal
supply cable to the mains socket with PE conductor. The current cable may only be attached or removed
when the device is switched off. The connecting socket for the power cable is located at the back of the
housing.
Foot switches are connected at "1mono/bi" or "2 mono/bi + auto". The terminals are located at the back
of the housing.
Handles and electrodes are connected correspondingly at the front of the housing to "MONOPOLAR1",
"MONOPOLAR2" or "BIPOLAR".
8.2 Connection of argon
The HBS x-touch 300 argon BiSeal
transmission system.
Gas flow from gas transmission system
Close the hose at one of the two connection points (D or G).
Use of gas cylinders
Secure the gas cylinder (A) with the provided fixture (B)
Do NOT remove the identification label
Mount an original HEBU pressure relief valve HEBU (C) at the
gas cylinder (article number: HF 9530-02)
50
Please observe the following warning remarks with regard to
argon-supported coagulation and cutting
Before first using the electrosurgical unit, we recommend
familiarizing employees with the operating characteristics of the
different operating modes and power settings by carrying out
practice exercises using fresh meat (beef, pork).
Use ultra-pure argon gas with a purity level of 99.996 % or
higher. Using gas with a lower purity can place the safety of
patients in jeopardy.
®
is able to work both with gas from a cylinder or from a gas
HBS x-touch 300 argon BiSeal
®
electrosurgery unit is connected via the
®