4. Safety functions
4.1. Lock button
(Only for electrically adjustable versions).
The lock button is located in the middle of the undercarriage.
Once the lock button has been depressed all electrical func-
tioning will stop. Rotate the lock button a quarter of a turn to
the right to unlock it (See figure 12).
The orange LED light on the manual control will flash
when the Lock button has been activated.
4.2. Sleep setting
After around an hour, the stretcher will fall into a 'sleep' setting and the battery indica-
tion LEDs on the manual control will go out. In order to use the stretcher again, it must
be activated (See paragraph 2.2. Activating).
4.3. Emergency lowering
(Only for electrically adjustable versions).
If the controls stop working and the stretcher is in a raised
position, it can be lowered using the emergency lowering
button (See figure 13).
Check first to make sure the stretcher is not in sleep setting!
• Remove the plug from the manual control (1).
• Press the emergency lowering button (2).
The stretcher will now return to the lowest position (make sure no persons or objects are
under the stretcher). Take the client from the stretcher and make sure the device is not
used until it has been repaired. Place the stretcher in a safe place and indicate that it is
out of order.
Contact the technical service or Lopital's service department.
E-mail: service@lopital.nl
4.4. On / Off switch
(Only for electrically adjustable versions).
In the centre of the undercarriage, there is an on / off
switch. If the shower stretcher is not used for an extended
period, you are advised.
(Only for electrically adjustable versions).
Marina Shower stretcher | Instructions for use | 39
Figure 12 Lock button.
1
Figure 13
Emergency lowering button.
Figure 14 On / off switch.
2