Installation
RotoForce-4 on RotoPol-31
RotoForce-4
Instruction Manual
Lift off the cap covering the supporting hole placed on the top
of the RotoPol-31 cabinet with a knife.
Likewise remove the two caps covering the fastening screws on
the left-hand side of the RotoPol cabinet.
Remove the black stop ring from the supporting column of
RotoForce-4.
Lead the supporting column of RotoForce-4 down into the
supporting hole in RotoPol-31. Turn the column until the cross
pin in the bottom of the column gets caught by the guiding
slots inside RotoPol-31.
Tighten the two fastening screws on the left-hand side of the
RotoPol-31 cabinet and cover the screws again with the two
caps.
Adjust the height of RotoForce-4 in relation to the preparation
disc:
Mount a preparation disc on RotoPol-31.
Mount a specimen mover plate in the coupling of
RotoForce-4.
Lower RotoForce-4 into working position (see The Handle
in Section Operation).
Measure the distance between the specimen mover plate
and the preparation disc. The correct distance is 1 -
1.5 mm.
If the distance is not correct, an adjustment has to be made:
Move RotoForce-4 into upper position and secure it there
(see The Handle in Section Operation)
Disengage the stop ring by loosening the retaining screw.
Move the stop ring into correct position and fasten the
retaining screw again.
Lower RotoForce-4 into working position and measure
the distance again. If it is still not correct, repeat the last
3 steps until the correct distance is obtained.