Installation
RotoForce-4 on RotoPol-31
RotoForce-4
Instruction Manual
Lift off the cap covering the supporting hole placed on the top
of the RotoPol-31 cabinet with a knife.
Likewise remove the two caps covering the fastening screws on
the left-hand side of the RotoPol cabinet.
Remove the black stop ring from the supporting column of
RotoForce-4.
Lead the supporting column of RotoForce-4 down into the
supporting hole in RotoPol-31. Turn the column until the cross
pin in the bottom of the column gets caught by the guiding
slots inside RotoPol-31.
Tighten the two fastening screws on the left-hand side of the
RotoPol-31 cabinet and cover the screws again with the two
caps.
Adjust the height of RotoForce-4 in relation to the preparation
disc:
−
Mount a preparation disc on RotoPol-31.
−
Mount a specimen mover plate in the coupling of
RotoForce-4.
−
Lower RotoForce-4 into working position (see The Handle
in Section Operation).
−
Measure the distance between the specimen mover plate
and the preparation disc. The correct distance is 1 -
1.5 mm.
If the distance is not correct, an adjustment has to be made:
−
Move RotoForce-4 into upper position and secure it there
(see The Handle in Section Operation)
−
Disengage the stop ring by loosening the retaining screw.
−
Move the stop ring into correct position and fasten the
retaining screw again.
−
Lower RotoForce-4 into working position and measure
the distance again. If it is still not correct, repeat the last
3 steps until the correct distance is obtained.
3