5. Maintenance
Clean all accessible surfaces of the machine with a moist cloth.
Raise RotoForce-4 to its upper position.
Select SAMPLE HOLDER and press START
will move down without rotating. Clean the coupling with a
clean cloth. An acid-free oil is recommended for the lubrication
of the coupling to ensure easy movement.
o
Press STOP
to return the coupling into the cylinder.
Select SINGLE SAMPLES and press START
feet will move down. Clean the pressure feet with a clean
cloth. An acid-free oil is recommended for lubrication of the
pressure feet, to ensure easy movement.
o
Press STOP
to return the pressure feet into the cylinder.
To avoid clogging of the Lupo nozzle, you should clean the nozzle
once a week:
Set the mode switch on Lupo to STOP
Remove Lupo from the supporting arm.
Hold the Lupo lubricant bottle firmly and remove the top lid.
Empty the bottle, still mounted on Lupo, of any remaining
lubricant.
Fill the bottle with mild, soapy water.
Put Lupo back on the supporting arm.
Set the mode switch on Lupo to START
Set the dosing control to max. dosing (step 10) and let the
water run through the nozzle for 2-3-min.
Repeat the above points but fill with tap water to rinse
out any soap residue.
Refill with lubricant and remount the top lid.
s