Inserting a Specimen Holder/
Specimen Mover Plate
Setting Holder Type
Adjusting the Force
Adjusting the Preparation Time
Adjusting the Position of
RotoForce-4
Running the preparation
26
RotoForce-4
Instruction Manual
Quick Reference Card
Place the specimen holder or specimen mover plate under the
quick coupling.
Press and hold the release button on the column of RotoForce-
4, while guiding the pressure tap of the specimen holder or
specimen mover plate into the coupling.
Turn the specimen holder or specimen mover plate until the
three pins engage in the corresponding holes.
Release the button.
m
Press MENU
L
Press SELECT
SINGLE SAMPLES.
fO
Press FORCE
value.
tO
Press TIME
value.
Unlock RotoForce-4 by moving the handle up to two dots.
Adjust the RotoForce-4 so that the specimens are running at
about 2 - 4 mm from the periphery of the preparation disc.
Lock RotoForce-4 in that position by moving the handle down.
Adjust time and force on the RotoForce-4.
Set RotoForce-4 to Auto.
Set Lupo to Auto if lubricant dosing is required.
s
Press START
on the RotoForce-4 to reach the menu system.
to toggle between SAMPLE HOLDER or
fN
or
to adjust the force to the desired
tN
or
to adjust the time to the desired
on RotoPol.