Step 13
Step 14
Step 15
Remove the
blanking plug from
the test port.
CAUTION
Do
not remove the
blanking plug
when the
instrument is
Step 16
under pressure.
Before removal
of the blanking
plug, look at the
instrument
pressure screen
for pressure. If
pressure is
seen, turn the
pressure release
knob fully
counter-
clockwise until
the pressure
Step 17
screen shows
zero (gauge
sensor) or
atmospheric
pressure
(absolute
sensor).
2.3.3 Supply pressure
INFORMATION
Turn the pressure
release knob fully
before attempting to generate pressure.
clockwise until it is
Note: Make sure the used pressure limits are in the limits
hand-tight.
given on the data label on the instrument bottom.
Turn the volume
adjuster wheel 2 to 5
turns clockwise to
remove any caught
air.
Note: There can
be air bubbles at
the opening of the
test port.
CAUTION
Do
not operate the
priming pump.
Make sure the instrument is primed
Step 1
Copyright 2022 Baker Hughes Company.
English–DPI 610E Quick Start & Safety Manual | 9
Make sure the
device under test is
set in position and
connect the device
under test to the test
port (turn the
connector counter-
clockwise).
Note: Use the
adapters supplied
with the
instrument or an
AMC adapter(s)
and the related
seal(s).
Set the device under
test into position.
Use the priming
pump to prime the
system to a
maximum of 10-20
bar.
After the instrument is
primed, turn the volume
adjuster wheel
clockwise until the
necessary pressure is
achieved and let the
pressure become
stable.
Note: If the
necessary pressure
is not supplied or not
stable, release the
pressure in the
instrument and do
the priming sequence
again.