Step 2
Step 3
Step 4
Step 5
Step 6
Step 7
Copyright 2022 Baker Hughes Company.
8 | DPI 610E Quick Start & Safety Manual–English
Connect the
reservoir to the
reservoir port and fill
the reservoir to the
correct level. See
Chapter 2.3.1 on
page 7
Energize the
instrument. Push
down and keep
down the power
button for 2
seconds. Wait for
the display to show
the Druck splash
screen.
Step 10
Select Normal from
the Channel
Function screen.
From the dashboard
select:
CALIBRATOR >>
or
>>
FUNCTION >>
Step 11
Pressure >> INT >>
Normal
Select the wanted
measurement units.
From the dashboard
select:
CALIBRATOR >>
or
>>
UNITS
Step 12
Turn the volume
adjuster wheel
clockwise until it
stops.
Turn the pressure
release knob fully
clockwise until it is
hand-tight.
Step 8
Step 9
Slowly operate the
priming pump until
the pressure
reaches 5-10 bar.
Turn the volume
adjuster wheel fully
counter-clockwise
until it stops.
Slowly operate the
priming pump until
the pressure
reaches 5-10 bar.
Turn the pressure
release knob a
quarter of a turn
anti-clockwise to
release the
pressure.
Slowly operate the
priming pump until
no air bubbles can
be seen in the
reservoir.
Note: It will take
10-15 strokes to
remove caught air
from the system.
CAUTION
Do
not operate the
pump when no
air bubbles can
be seen in the
reservoir.