Les langues disponibles

5.
It is strictly prohibited to use strong chemical solvent to clean the surface.
6.
Please clean the scale with dry or wet dishcloth. Washing it with water is strictly-
prohibited.
7.
Please don't use the same power supply circuit together with the big power
equipment, because the running of the big power equipment will influence the
weighing.
CAUTION
If the weight window displays "—Lb—" when the weight is zero, it means the
voltage of the accumulator is low. Under this condition, you can use the scale in
a short time, but you should plug in the AC plug as soon as possible to recharge
the accumulator, please recharge the accumulator for more than 12 hours.
The accumulator is not included in free maintain warrant.
SBS-ZW 3001H Counting Scale Calibration procedure(New program)
During self checking,press "[Sample]+[8]" key at the same time,after the self
check,the three windows will display as:[Unit],[g], [......] can use [Tare] key select
the unit (in here have Kg and g available). press the number key input"090316"press
[ACCU] , key enter to set process.
EN
1.
"Capacity choose". three windows display as: [FULL], [SCALE], [XXX](xxx mean
the capacity)can use number key input the capacity (example:30Kg scale,that
input 30) if unit is g,that input 30000). Drücken Sie die Taste [ACCU] zur Bestäti-
gung.
2.
Division choose,three windows display as: [......], [d-g], [xx] (x.x steht für die Ein-
teilung). (x.x means the div.) press [TARE] key choose you want the division (in
there have 1/3000-1/600000 available) press[ACCU] key confirm it.
3.
"Zero track range choosen",the three windows shows as: [......], [ZERO], [1.5]
press [TARE] key (in there has 0.5 1.5 2.5 3.5 available) if you choose the value
bigger,the zero track rang will bigger),press [ACCU] zur Bestätigung.
4.
"Calibration mode choosen", press [TARE] key choosen calibration mode. in here
has A: "Calibration by any weight, B: "calibration by two point" and C: "calibration
by three point", Operation as below:
A, [1], [LOAD], [......] for calibration by any weight,press [ACCU] keyconfirm,display
the capacity value,at this time can input the weight value by keyboard,
then put the weight same just input that weight value on the pan.press [ACCU] key
when display is stable and on the weight windows show the number same the weight
value.The calibration finish.
Note: when you select the calibration by any weight,the weight value
couldn't smaller than 1/3 of full capacity.
12
B: [2], [HF-FLUL], [......] for calibration two point, Press [ACCU], key showAD
inner code,when press [ACCU] key show "HALF",put on1/2 full of capacity, wait the
number stable,press [ACCU] key again ,will show "FULL",please put on full of capaci-
ty and stalbe after,press [ACCU] key confirm it.the scale will display the weight value
and the calibration finished it.
C: [3], [LOAD3], [......] For calibration three point.press [ACCU] , key display
AD inner code,press [ACCU] key again display "LOAD1", put on 1/3 full of capacity
weight,press [ACCU] key confirm,display "LOAD2",this time put on 2/3 full of capaci-
ty weight, press [ACCU] key confirm,when display "LOAD3",put on full of capacity
weightpress [ACCU] key confirm.the scale will display the weight value and the calib-
ration finished it.
Note:during the set,can press [AC] key return back to up menu.
Simple calibration:
During any weighing,press [AC] key about 3seconds,that three windows will show as:
[0] [LOAD], [current capacity] input calibration weight value by number key and put
on the weight, when the stable sign appear it, press [ACCU] key confirm,the calibrati-
on is finished.