PINKWAVE™ OPERATION
Selecting Your Curing Mode
1. The curing modes are selected via the button on the top of the device.
2. Toggle through modes by pressing the mode button.
3. The current mode will light up bright pink when it has been selected.
This setting will remain the default setting until a new setting is selected.
4. On standard and ramp mode, the light will beep once at 5 seconds,
twice at 10 seconds and three times at 15 seconds.
The figure below illustrates the light position based on the current mode.
3B
20R
20S
= Mode selected
Quick Curing Guide
Irradiance (mW/cm
)
2
Per 2 mm Layer
Final Cure
NOTE:
Curing time may need to be adjusted due to composite, shade, depth of layer
if over 2mm.
NOTE:
These are default curing durations for the PinkWave™ Curing Light.
Follow composite / dental material manufacturer's guidelines for required
curing durations and irradiance values.
NOTE:
Place the PinkWave™ Curing Light as close as possible to dental material.
Irradiance rapidly decreases with increasing distance.
NOTE:
Software will automatically turn the PinkWave™ handpiece off if left on
for two continuous minutes in white light / transilluminator mode.
NOTE:
Software does not allow for the handpiece to be used when in the charging base.
Transilluminator
20R
20S
Standard Mode
>1515
One 10s cure
One 20s cure
3B
3-second Boost Mode
20-second Ramp Mode
20-second Standard Mode
All
Transilluminator
NOTE: Setting is memorized as
default until charged.
Ramp Mode
>1515
One 20s cure
One 20s cure
Boost Mode
>1720
One 3s cure
Two 3s cures