IMMUNITY TEST
TEST LEVEL
Voltage dips, short
interruptions and
voltage variations
on power supply
0% UT; 1 cycle and 70% UT; 25/30 cycles
input line IEC
Single phase at 0° 0% UT; 250/300 cycles
61000-4-11
Power frequency
(50/60Hz) magnetic
field IEC 61000-4-8
Portable and mobile RF communications equipment should not be used closer to any part of PinkWave™, including cables,
than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Conducted RF
IEC 61000-4-6
3V from 0.15 to
Radiated RF
6V from 0.15 to
IEC 61000-4-3
80% AM at 1kHz
3V/m from 80MHz
Table: According to IEC 60601-1-2, Edition 4.0
NOTE: UT is the a.c. mains voltage prior to application of the test level.
NOTE: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a - Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered.
If the measured field strength in the location in which the PinkWave™ is used exceeds
the applicable RF compliance level above, the PinkWave™ should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the PinkWave™.
b - Over the frequency range 150 kHz to 80 MHz, field strength should be less than 10 V/m.
22
IEC 60601
COMPLIANCE
LEVEL
0% UT; 0.5 cycle at 0°, 45°, 90°,
135°,180°, 225°, 270° and 315°
30 A/m
30 A/m
3V, 6V
3Vrms, 6V
3V/m
3V/m
3V from 0.15 to
80MHz;
80MHz;
6V from 0.15 to
80MHz and
80MHz and
80% AM at 1kHz
3V/m from 80MHz
to 2.7GHz
to 2.7GHz
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
Mains power quality should be that of typical
commercial or dental environment. If the user of the
PinkWave™ requires continued operation during power
mains interruptions, it is recommended that
the PinkWave™ be powered from an uninterruptible
power supply or battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or dental environment.
Portable and mobile RF communications equipment
should be used no closer to any part of the PinkWave™
Curing Light, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
3.5
12
d = [
] P
d = [
] P
V
V
1
2
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m). Field strengths from fixed RF
transmitters, as determined by an electromagnetic site
survey, should be less than the compliance level in each
frequency range. D Interference may occur in the vicinity
of equipment marked with the following symbol:
12
23
d = [
] P
d = [
] P
E
E
1
1