REPROCESSING INSTRUCTIONS
INSTRUCTIONS
Preparation
before cleaning:
Cleaning &
Disinfection:
Automated
18
• Following clinical use, immediately wipe clean any visible
debris or contamination using a lint free cloth, which may be
dampened with deionized water.
• For non-electrical devices or accessories, the product should be
immediately rinsed with water to remove debris.
• For non-electrical devices or accessories, if reprocessing cannot
be performed immediately, soak product in a mild, pH-neutral
detergent until cleaning can be performed and maintain
separation of soiled product from non-contaminated devices
to avoid cross-contamination. DO NOT soak or submerge
electronic devices.
• If reprocessing cannot be performed immediately, maintain
separation of soiled product from non-contaminated devices to
avoid cross-contamination.
• No disassembly is required for reprocessing. Do not attempt to
disassemble product.
• Remove any visible gross debris and wipe using a
lint free cloth. For non-electronic devices or accessories,
water may also be used.
• For non-electronic devices and accessories, it is recommended
to use a washer-disinfector meeting the requirements of
ISO 15883. Products should be positioned such that they do
not make physical contact with other parts in the washer-
disinfector. Use a neutral-pH based cleaning agent. There are
no known constraints on pressures and temperatures.
NOTE: Electronic devices should never be placed in an automated
washer-disinfector and should never be submerged or exposed to
free flowing cleaning agents or water.
(continued)