REPROCESSING INSTRUCTIONS
Manufacturer:
Device:
WARNINGS
Limitations on
process:
PinkWave™ & Accessories
• Appropriate PPE (gloves, safety glasses, scrubs or lab coat)
should be worn when reprocessing medical devices.
• Electronic devices should never be submerged or soaked in any
cleaning solution.
• Electronic devices should never be autoclaved.
• Do not disassemble electronic devices.
• Do not use water on electronic devices.
• Do not use abrasive cleaners.
• Use neutral pH detergents during cleaning.
• Do not use with commercial sterilization wraps or pouches with
sterilization indicators that leave a stain.
• For autoclavable products, do not sterilize with other products.
Do not allow products to touch one another during autoclaving.
• Reprocess products as soon as possible after use.
• It is the user's responsibility to ensure that all reprocessing
equipment used is properly validated, maintained, and calibrated
in accordance with equipment specification and requirements.
• Reference should be made to the device's instructions for use /
user manual for additional information which may be applicable
to reprocessing.
• The products should be thoroughly inspected and tested
for functionality after each reprocessing cycle. Devices should
be replaced if any visible wear is seen, or if any noticeable
decrease in product functionality is observed. Operator use and
care will impact how long product will last. Contact Vista Apex
with any questions.
• Vista Apex has validated the following number of reprocessing
cycles for the products:
- PinkWave™ Silicone Shields & Anti-Roll Brackets: 80 cycles
Inter-Med, Inc.