APPENDIX A: EMC INFORMATION
If you have any questions regarding the guidance and declarations listed below, please
contact your home care provider.
Guidance and Manufacturer's Declaration - Electromagnetic Emissions: This device is
intended for use in the electromagnetic environment specifi ed below. The user of this
device should make sure that it is used in such an environment.
EMISSIONS TEST
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fl uctuations/Flicker
emissions IEC 61000-3-3
Emission of Radio Frequen-
cy Energy
RTCA/DO-160G Section 21
Guidance and Manufacturer's Declaration - Electromagnetic Immunity: This device is
intended for use in the electromagnetic environment specifi ed below. The user of this
device should make sure that it is used in such an environment.
IMMUNITY TEST
Electrostatic
±15kV air
Discharge (ESD)
±8kV contact
IEC 61000-4-2
Electrical Fast
±2kV for power
Transient/Burst
supply lines
IEC 61000-4-4
±1kV for input-output
lines
Surge
±1 kV di erential
IEC 61000-4-5
mode
±2 kV common
mode
COMPLIANCE
Group 1
Class B
Class A
Complies
Category M
IEC 60601 TEST
COMPLIANCE
LEVEL
±15kV air
±8kV contact
±2 kV for supply
mains
±1 kV for input/out-
put lines
±1 kV di erential
mode
±2 kV common
mode
11/167
ELECTROMAGNETIC
ENVIRONMENT GUIDANCE
This device uses RF energy
only for its internal function.
Therefore, its RF emissions
are very low and are not
likely to cause any interfe-
rence in nearby electronic
equipment.
The device is suitable for
use in all establishments,
including domestic estab-
lishments and those directly
connected to the public
low-voltage power supply
network.
This device is suitable for
use onboard commercial air-
planes inside the passenger
cabin.
ELECTROMAGNETIC
LEVEL
ENVIRONMENT
GUIDANCE
Floors should be wood,
concrete or ceramic
tile. If fl oors are co-
vered with synthetic
material, the relative
humidity should be at
least 30%.
Mains power quality
should be that of a ty-
pical home or hospital
environment.
Mains power quality
should be that of a ty-
pical home or hospital
environment.
EN