The fl oat in the fl ow meter
does not move up or down
when adjusting the fl ow
control dial.
There is water in the can-
nula.
EN
5. SPECIFICATIONS
Technical Specifi cations
Model
Model
Power consumption (W)
Rating current (A)
Electrical requirements
Flow rate (l/min)
Purity
Outlet pressure (mpa)
Dimensions
Net weight
Sound level
Equipment class and
type
EMC compliance
Features
Disposal
Dispose of this device in accordance with local regulations.
The fl ow control dial is not
open.
The fl ow control dial is
faulty.
There has been a change
in temperature;
OR
The device is too clo-
se to a wall, curtains or
furniture.
OC-E80
OC-E80
650
750W 6A (North America)
2.25
220V (+/- 10%) 50HZ (+/-10%)
220V (+/- 10%) 60HZ (+/-10%)
115V (+/- 10%) 60HZ (+/-10%)
0,5 to 8
87 – 96%
0.05±0.005
381 mm x 347 mm x 689 mm
27 kg
50 dBA or less
Class IIa
IEC 60601-1-2
Standard: LCD display; oxygen concentration indicator; tempe-
rature alarm; high/low pressure alarm; low oxygen concentra-
tion alarm; power failure alarm; timed setting; intelligent fault
diagnosis; maintenance reminder.
Additional: Positive pressure outlet
10/167
Turn the fl ow control dial slowly
and carefully.
Contact your home care provider.
Dry the inside of the humidifi er
cover.
Do not use hot water in the humi-
difi er bottle.
Do not overfi ll the humidifi er
bottle.
Keep the oxygen concentrator
and cannula in the same room at
the same temperature.
If using an extension hose with
your oxygen concentrator, speak
to your home care provider ab-
out fi tting a water trap to collect
excess moisture.
OC-E100
OC-E100
650
750W 6A (North America)
2.25
220V (+/- 10%) 50HZ (+/-10%)
220V (+/- 10%) 60HZ (+/-10%)
115V (+/- 10%) 60HZ (+/-10%)
1 to 10
87 – 96%
0.05±0.005
381 mm x 347 mm x 689 mm
27 kg
50 dBA or less
Class IIa
IEC 60601-1-2