7. Calibration

Calibration of the climate sensors is only necessary in rare cases. If high
demands on the accuracy are made, we recommend an annual one-
point calibration of the sensor.
You can perform one-point calibration yourself but we advise against
doing so because professional reference values are not usually
available.
Instead, make use of the possibility of certified calibration according to
DKD and ISO. Further information about this can be obtained from
your dealer.
One-point calibration (r.H.) with calibration block and
calibration ampoule:
Sensor
head
Calibration
ampoule
Operating manual – English
Check the calibration block for foreign
bodies or residue from previous cali-
brations and clean the block if neces-
sary.
Calibration
Different calibration fluids are available
block
for the moisture values 35 %, 50 % and
80 %. Only the 50 % calibration fluid
should be used for standard calibrations.