One-point calibration of temperature
and relative humidity
With CAL a one-point calibration can be carried out for the sensor
displays of temperature (CAL 1) and relative humidity (CAL 2).
All sensors are already calibrated at the factory and have an appropriate
factory calibration characteristic.
In the one-point calibration a global calibration curve shift is made by
specifying an offset which is effective over the whole measuring range.
The offset to be entered is the value by which the calibration curve is
shifted.
TH
IR
Operating manual – English
CAL 1: The offset for sensor 1 (temperature)
is set with CAL 1 (one-point calibration).
The offset is shown in the lower display.
A maximum ± 10 °C or ± 10 °F can be set.
°C
Selection can be made with
confirmation with
°C
The factory settings are obtained by
CAL 1
setting the offset to 0.0.
DP
CFG
and
,
.
B - 14