WARNING
118
The apparatus shall not be exposed to dripping or
splashing. No objects filled with liquids, such as vases,
shall be placed on the apparatus.
WARNING
Make sure the surface is wide enough so that this
apparatus's width and depth don't exceed the surface's
edges.
If not, this apparatus may lean or fall over and cause an
injury.
WARNING
To prevent injury, if mounting the unit using a mounting
arm, wall fixture, or other mounting device prepared by
the customer, mount the unit securely as described in the
instruction manual provided with the mounting device.
Check beforehand that the mounting device used has
sufficient strength to support the added weight of the
unit.
Check yearly that the mounting device is securely
attached.
124
Consult with Sony qualified personnel for the following
types of installation location.
– Wall mount
– Ceiling mount
– Mounting arm
Caution PARALLEL REMOTE connector
(modular connector, 8-pin)
• Do not come into contact with this connector and
patients at the same time.
Doing so may result in a generation of voltage that can
be harmful to patients if the unit is malfunctioning.
Always disconnect the power cord before connecting
and disconnecting connectors.
• For safety, do not connect the connector to peripheral
device wiring that might have excessive voltage.
Follow the instructions for use for this port.
Caution
SERIAL REMOTE connector
(RJ-45)
113
• Do not come into contact with this connector and
patients at the same time.
Doing so may result in a generation of voltage that can
be harmful to patients if the unit is malfunctioning.
Always disconnect the power cord before connecting
and disconnecting connectors.
• For safety, do not connect the connector to peripheral
device wiring that might have excessive voltage.
Follow the instructions for use for this port.
8
GB
123
113
76
76
WARNING
27
Using this unit for medical purposes
The connectors on this unit are not isolated.
Do not connect any device other than one which
conforms to IEC 60601-1 standards.
When an information technology device or AV device
that uses an alternating current is connected, current
leakage may result in an electric shock to the patient or
operator.
If use of such a device is unavoidable, isolate its power
supply by connecting an isolation transformer, or by
connecting an isolator between the connecting cables.
After implementing these measures, confirm that the
reduced risk now conforms to IEC 60601-1 standards.
WARNING
115
For the DC power supply, make sure to use the supplied
AC adapter, AC-110MD.
If another power supply is used, there is a risk of fire or
electric shock.
Caution
121
The unit is not disconnected from the AC power source
(mains) as long as it is connected to the wall outlet, even
if the unit itself has been turned off.
Caution
157
Do not use the device in a MR (Magnetic Resonance)
environment.
It may cause a malfunction, fire, and unwanted
movement.
For details on operation, refer to the manuals
stored on the supplied CD-ROM.
Using the CD-ROM manuals
The manuals can be viewed on a computer with Adobe
Reader installed.
You can download Adobe Reader for free from the
Adobe website.
1. Open the "index.html" file on the CD-ROM.
2. Select the language of the manual you want to view.
Note
If you damage or lose the CD-ROM, you can purchase
a new one from your dealer or Sony service
representative.
163-2

Publicité

loading