16.4 Measurement tolerances
Accuracy in ± (% of reading + display error in counts (= number of smallest
points)). These accuracy readings are valid for one year at a temperature of
+23 °C (± 5 °C) and relative humidity of less than 80 % (non-condensing). If the
multimeter is used outside of this temperature range, use the following coefficient
to calculate the accuracy. +0.1 x (specified accuracy)/1 °C
The accuracy of measurements may be affected when the device is used in a high-
frequency electromagnetic field.
Direct voltage (V/DC)
Range
60.00 mV*
600.0 mV*
6.000 V
60.00 V
600.0 V
1000 V
*Only available in "mV" mode
Specified measurement range: 5–100 % of the measurement range
1000 V overload protection; impedance: 10 MΩ
The multimeter may display ≤10 counts if a measurement input is short-
circuited.
The LoZ low impedance measurement is not specified.
84
Resolution
0.01 mV
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
Accuracy
±(0.5% + 10)
±(0.5% + 5)
±(0.5% + 5)
±(0.5% + 5)
±(0.5% + 5)
±(0.8% + 5)