GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC IMMUNITY
The Apica Air is intended for use in the electromagnetic environment specified below. The customer or
the user of the Apica Air should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic
+/- 8 kV
discharge (ESD)
contact
IEC 61000-4-2
+/- 2 kV, +/- 4 kV,
+/- 8 kV, +/- 15 kV
air
Electrical fast
±2kV
transients/bursts
100kHz
IEC 61000-4-4
repetition frequency
Surge
Line to line:
IEC 61000-4-5
±0.5kV,
±1kV
Line to earth:
±0.5kV,
±1kV, ±2kV
Voltage dips
0% UT; 0.5 cycle at 0°,
IEC 61000-4-11
45°, 90°, 135°, 180°,
225°, 270°, and 315°
Voltage
0% UT; 1 cycle and
interruptions
70% UT; 25/30 cycles
IEC 61000-4-11
sine phase at 0°
0% UT; 250/300 cycle
Rated Power
30 A/m 50Hz or 60Hz
frequency
magnetic field IEC
61000-4-8
NOTA: UT: rated voltage(s); E.g. 25/30 cycles means 25 cycles at 50Hz or 30 cycles at 60Hz
Compliance level
+/- 8 kV
contact
+/- 2 kV, +/- 4 kV,
+/- 8 kV, +/- 15 kV
air
±2kV
100kHz
repetition frequency
Line to line:
±0.5kV, ±1kV
Line to earth:
±0.5kV,
±1kV, ±2kV
0% UT; 0.5 cycle at 0°,
45°, 90°, 135°, 180°,
225°, 270°, and 315°
0% UT; 1 cycle and
70% UT; 25/30 cycles
sine phase at 0°
0% UT; 250/300 cycle
30 A/m 50Hz or 60Hz
Electromagnetic environment
- guidance
Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of devices require continued
operation during power mains
interruptions, it is recommended
that devices be powered form
an uninterruptible power supply
or a battery
Power frequency magnetic field
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.