GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC IMMUNITY
The Apica Air is intended for use in the electromagnetic environment specified below. The customer or
the user of the Apica Air should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted
3V
dis- turbances
0.15 MHz – 80 MHz, 6
induced
V in ISM bands
by RF fields
between 0.15 MHz and
IEC 61000-4-6
80 MHz,
80 % AM at 1 kHz
Radiated
3 V/m,
RF EM fields
80 MHz – 2,7 GHz,
IEC 61000-4-3
80 % AM at 1 kHz
Proximity fields
See the RF wireless
from RF wireless
communicati on
communication
equipment table in
equipment
"Recommen ded
IEC 61000-4-3
minimum separation
distances"
44
Compliance level
3 V
3 V/m
Complies
Electromagnetic environment
- guidance
Portable and mobile RF
communications equipment
should be usedno closer to any
part of the Apica Air, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended minimum
separation distances
See the RF wireless
communication equipment
table in "Recommended
minimum separation distances"