7. MAINTENANCE

File clip
Note
File clip and lip hook must be pre-cleaned and sterilized before every use.
• Pre-cleaning: Use a soft brush or moist cloth to remove visible contaminations, then wipe all the
surfaces with a cloth lightly moistened with Ethanol for disinfection (Ethanol 70 to 80vol%) at least 1
min, repeat for 5 times.
• Packing: Pack each component in a separate steam-sterilization pouch.
• Sterilization: Steam sterilization at 134°C at least 6 minutes .
Minimum drying time after sterilization: 10 minutes.
• Storage: Keep the components in sterilization packaging in a dry and clean environment.
Warning
• Comply with your national guidelines, standards and requirements for cleaning, disinfection
and sterilization.
• Be careful to avoid cross contamination when performing maintenance.
• Must be autoclaved after use for each.