RECOMMENDED MINIMUM SEPARATION DISTANCES
Nowadays, many RF wireless equipments have being used in various healthcare locations where medical
equipment and/or systems are used. When they are used in close proximity to medical equipment and/or
systems, the medical equipment and/or systems' basic safety and essential performance may be affected.
The Apica Air has been tested with the immunity test level in the below table and meet the related
requirements of IEC 60601-1-2:2014. The customer and/or user should help keep a minimum distance
between RF wireless communications equipments and the Apica Air as recommended below.
Test
Band
frequency
(MHz)
(MHz)
385
380-390
450
430-470
710
704 - 787
745
780
810
800-960
870
930
1720
1700-1990 GSM 1800; CDMA
1845
1970
2450
2400-2570 Bluetooth WLAN,
5240
5100-5800 WLAN 802.11 a/n
5500
5785
Service
TETRA 400
GMRS 460 FRS 460
LTE Band 13,17
GSM 800/90, TETRA
800, iDEN 820,
CDMA 850, LTE
Band 5
1900; GSM 1900;
DECT;
LTE Band 1,3,4, 25;
UMTS
802.11 b/g/n,
RFID 2450,
LTE Band 7
Max
Modulation
power
(W)
Pulse modulation
1.8
18Hz
FM
2
± 5 kHz deviation
1 kHz sine
Pulse modulation
0.2
217Hz
Pulse modulation
2
18Hz
Pulse modulation
2
217Hz
Pulse modulation
2
217Hz
Pulse modulation
0.2
217Hz
Immunity
Distance
test level
(m)
(V/m)
0.3
27
0.3
28
0.3
9
0.3
28
0.3
28
0.3
28
0.3
9
45