6.2 Function checking of APEX locator
After turning on, insert the tester into the APEX locator.
Clamp the groove of tester with file clip.
The measuring bar on the screen flashes at point 02, 03 or 04.
RecommendtotesttheAPEX locator with tester once a week.
Note
If the measurements are not expected, check that the tester is connected
properly. If the normal connection still does not show the
expected value, please stop using the device and contact the local dealer
for processing.
Confirm the device with short touch file clip and lip hook before.
Confirm that the tester is not installed on the APEX locator. Then connect
the file clip, lip hook and extension cord according to chapter 5.1.
Finally, touch the lip hook with exposed metal position on file clip, the
measurement on the screen should be shown as - 2.
Warning
If the measurement shown is not - 2, check that the connection is normal.
If the normal connection still does not show the expected value, please
stop using the device and contact the local dealer for processing.
33