Press ENTER once more to select the parameter shown below, and to set both related limit values.
Press this key
to select the digit you want to change, and then ENTER
Press this key
to move the cursor to the other limit value, and proceed as before.
Press this key
to select
Proceed in the same way to choose each parameter you want to measure, then set all related tolerance
values.
δ
When RPc ,PPc, R
C, or P
because these parameters include a higher number of setting options.
Example
So that the RPc parameter can be appropriately calculated, you need to set the range that will take
peaks into account (with active inch unit system). You may also select the type of limit values along
with the extent of the calculation range.
Selecting printing parameters
17.2
To access this menu option, select
Press this key
to move from an option to another. Only one of the first two options can be enabled.
Press ENTER
to enable/disable any other key. Once completed, press this key
menu, or this one
again to return to main menu.
Note
If you want to check a disabled parameter as you've just finished up measuring, simply enable it. No
need to take a new measurement. The parameter Rmr%, which matches the bearing ratio, is graphically
printed out as a Abbott curve.
Selecting used standard
17.3
To access this menu option, select Standard, and then validate through ENTER
Move the cursor
to desired standard that will be enabled at once. As next step, press this key
return to sub-menu Standard. Press
Note
Selecting ISO Standard 12085 with active inch unit system causes the gauge unit to display a message
telling you that this standard requires the metric system for the needed calculation, before switching
over automatically to that one.
6999.006.1503
Save
or Cancel. Validate through ENTER
δ
C are selected and further enabled, the tolerance screen looks different
Print
Settings, and then validate over ENTER
again to return to main menu.
Page 14/22
to set correct value.
and return to previous menu.
to return to sub-
A. Mathier / March 15
to