8. FUNCTION CHECK
9. TROUBLESHOOTING
Problem
No weight displayed
with a load on the scale.
0.00 does not appear
before weighing.
One segment is illumi-
nated either continu-
ously or not at all.
The display
appears.
The display "bAtt"
appears.
►
Perform a function check before each use.
A complete function check includes:
• a visual inspection for mechanical damage
• a visual inspection and function check of the display
• function check of all the controls shown in the
section entitled "Overview" on page 71
• a function check of the optional accessories
If you find faults or deviations during the function
check, as a first step try to remedy the fault with the aid
of the section entitled "Troubleshooting" from
page 109.
CAUTION!
Bodily injury
If you find faults or deviations during the func-
tion check which you are unable to remedy with
the aid of the section entitled "Troubleshooting"
from page 109, you may not use the device.
►
Have servicing and repairs carried out
exclusively by seca service or by an
authorized service partner.
►
Please see the section entitled "Servicing" on
page 111.
Cause / Solution
The scale has no power supply.
- Check whether the scale is switched on
A load was already on the scale before it was
switched on.
- Remove the load from the scale
- Switch the scale off and back on again
The corresponding point has a fault.
- Inform the service department
Battery block power is running down.
- Recharge battery block as soon as possible
Battery block discharged.
- Charge battery block
English
Function check • 109