9. D I S A S S E M B LY A N D A S S E M B LY O F T H E
M I G H T Y M E D I C C O O L I N G U N I T
Remove the large and small Base Seal
Rings. Lever out the large Seal Ring
from the recess in the notch using the
Filling Chamber Tool.
Removal of the large Base Seal Ring
Press the upper Screen with the Filling
Chamber Tool from the top to the bot-
tom out of the Cooling Unit base.
Pressing out the upper Screen
Reassemble in reverse order.
When assembling, make sure
that Seal Rings are properly
seated. The Mouthpiece Seal
Ring, especially, must sit com-
pletely in the cap and be no
longer visible.
The upper Screen must be inserted
carefully into the groove provided. Dis-
regarding this may allow plant parti-
cles to enter the Cooling Unit, so block-
ing this or be inhaled.
When using the Dosing Capsules dur-
ing operation, the Lower Screen will
normally not get polluted. If the screen
does have to be exchanged one day,
proceed as follows:
Remove the lower Filling Chamber
Screen using the Filling Chamber
Tool by pushing in along the indenta-
tion of the Filling Chamber. In this way
the Screen arches and detaches itself
from the notch.
Inserting a new Screen: It is easier to
insert the Screen if it was previously
curved and inserted with the curved
side facing up. Using the Filling Cham-
ber Tool, you can press the screen flat
so that the edges of the screen seat
into the groove at the bottom of the
Filling Chamber.
88