Télécharger Imprimer la page

Vapormed Mighty Medic Mode D'emploi page 295

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 104
10. D E M I G H T Y M E D I C V E R D A M P E R
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The MIGHTY MEDIC Vaporizer is intended for use in the electromagnetic en-
vironment specified below. The customer or the user of the MIGHTY MEDIC
Vaporizer should assure that it is used in such an environment.
Immunity
test
Electrostatic
± 6 kV contact
discharge (ESD)
discharge
IEC 61000-4-2
discharge
Electrical fast
± 2 kV for power
transient/burst
supply lines
IEC 61000-4-4
± 1 kV for input/
output lines
Surge
± 1 kV line(s) to
IEC 61000-4-5
line(s)
± 2 kV line(s) to
earth
IEC 60601
Compliance level Electromagnetic environ-
test level
± 6 kV contact
discharge
± 8 kV air
± 8 kV air
discharge
± 2 kV for power
supply lines
not applicable:
input/output
lines are
non-existent
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Floors should be wood,
concrete or ceramic
tile. If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Mains power quality
should be that of a
typical residential,
commercial or hospital
environment.
Mains power quality
should be that of a
typical residential,
commercial or hospital
environment.