9. D I S A S S E M B LY A N D A S S E M B LY O F T H E
M I G H T Y M E D I C C O O L I N G U N I T
Turn the Cooling Unit with a 90° ro-
tation counterclockwise in order to
remove it from the MIGHT Y MEDIC
Vaporizer.
Pull out the Mouthpiece from the Cool-
ing Unit Cap, rotate slightly back and
forth while doing this, and pull off the
Mouthpiece Seal Ring.
Pulling the Mouthpiece from the Cooling Unit
Cap
Open the Cap Lock by pulling it
rearwards.
Opening the Cap Lock
Remove the Cap Lock from the cap by
pulling up at one side.
Only if you intend to immerse
the Cooling Unit into alcohol
(ethanol) is it necessary to re-
move the Cap Lock. Extended contact
with alcohol can lead to whitening of
the Cap Lock label or disintegration of
the glue.
Removal of the Cap Lock
Separate the Cooling Unit Cap from
the Cooling Unit Base by pulling off the
cap upwards.
Pull the Cooling Unit Cap from the Cooling
Unit Base