Immunity
test
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Note: Uτ is the a.c. mains voltage prior to application of the test level.
IEC 60601
test level
<5% Uτ (>95%
dip in Uτ) for
½ cycle
40% Uτ (60% dip
in Uτ) for 5
cycles
70% Uτ (30% dip
in Uτ) for
25 cycles
<5% Uτ
(>95% dip in
Uτ) for 5 s
Compliance level Electromagnetic environ-
<5% Uτ (>95%
dip in Uτ) for
½ cycle
40% Uτ (60% dip
in Uτ) for 5
cycles
70% Uτ (30% dip
in Uτ) for
25 cycles
<5% Uτ
(>95% dip in
Uτ) for 5 s
296
ment – guidance
Mains power quality
should be that of a
typical residential,
commercial or hospital
environment. The
battery powered
MIGHTY MEDIC
Vaporizer is able to offer
continuing functionality
even if the energy supply
is interrupted.