4. Acquiring CEPH Images
Frankfurt plane
4.2.3
Nasal positioner
38
9. Click the Ready button after the patient has been positioned. No X-Ray
will be emitted at this point. The CEPH sensor and the 2
capturing mode.
10. Proceed to section 4.3. Initiating X-Ray Exposure.
SMV Mode
Follow the procedure below to prepare and position the patient.
7. The ear rods should fit properly along
the external auditory meatus of the
patient. The patient's Frankfurt Plane
should be parallel to the floor.
8. Ask the patient to swallow saliva and
remain still with his/her mouth closed
until X-Ray exposure is complete.
nd
1. Widen the distance between the two
ear rods.
2. The nasal positioner should be flipped
to the side and up to prevent it from
obstructing the image acquisition.
PaX-i (PCH-2500) User Manual
collimator will enter