If necessary, further adjustments can be made using the arrows to the right of
each number.
Please wait for a moment, while the rotating moves to its
initial scanning position.
NOTE
When you click the Confirm button,
■ The Ready button will begin blinking to show that it has been activated.
This means that the equipment is ready for X-ray exposure.
■ The rotating unit will move to its initial scanning position.
■ Three laser beams, Mid-sagittal plane, Frankfurt plane and Canine, will
be activated to facilitate patient positioning. These beams will disappear
after 20 minutes or if the Ready button is clicked.
The laser beam on/off icons are located:
- In the upper right side of the window:
NOTE
- On the Handle Frame:
10. Guide the patient to the equipment. Position the patient within the
equipment. For further information about patient positioning, refer to section 3.2.
Positioning the Patient.
3. Acquiring PANO Images
8. A default value for the tube voltage
(kVp) and current (mA) will be displayed
based on the gender and X-ray intensity.
9. Click the Confirm button for these
parameters to take effect.
■ The scan time and estimated DAP
(Dose Area Product) values will be
shown on the main display for the
selected exposure.
15