Hardware Setup

Site of Operation

The site of operation must meet the temperature, humidity, and
atmospheric pressure requirements listed on page 22. In addition,
observe the following guidelines:
!
Make sure that the platform which supports the module is
large enough, level, and stable.
!
Make sure that the fan exhaust screen at the rear of the mod-
ule and the ventilation holes on the underside are not
obstructed.
!
Place the module at least 25 cm (10 inches) from any possible
source of ignition, such as sparking.
!
Place the module close enough to the patient so that the sam-
pling tubing can reach the airway T-connector and the
exhaust tubing the hospital's exhaust gas scavenging system
without stretching.
WA RN IN G S :
!
!
!
!
!
!
Page 4
Do not use mobile phones within 33 feet (10 m) of the
Scio module. Wireless phones may cause failure.
Do not expose the Scio Module to mechanical vibra-
tions or shock during measurement. Mechanical vibra-
tions or shock can have adverse effects on gas
measurement values.
The operation of the Scio module in magnetic reso-
nance imaging environments (MRI) is not supported.
Do not use a Scio module near devices with microwave
or other high-frequency emissions. These emissions
may interfere with the modules' operation.
SC 7000/8000/9000XL and the Scio Module must both be
powered from hospital grade connectors within the
same medically used room.
When the SC 7000/8000/9000XL is used with the Scio it
meets the Class A limits of CISPR11. The system is not
intended for connection to public mains.
Scio MODULE D011100
SC 7/8/9XL VF2.x