Message
Condition
MultiGas H/W
Loss of communication or hardware
failure
problem detected
MultiGas
Scio H/W or S/W incompatibility
Incompatible
MultiGas
Device initializing
Initialization
MultiGas Oxygen
Oxygen calibration successful
Cal. Accepted
MultiGas Oxygen
Oxygen calibration in progress
Cal. in Progress
MultiGas Sample
Scio module sample line occluded
Line Occlusion
Watertrap full, defective, or not
installed.
MultiGas
Scio module has become discon-
Unplugged
nected
MultiGas Zero in
Scio module zero in progress
Progress
Accessories
Scio Water Trap (12 pieces)
Scio Sample line (Set of 10)
Power supply
SC 7/8/9XL VF2.x
Suggested Action
• Check connections.
• Unplug and re-plug Scio module.
• Power-cycle monitor or un-dock
and re-dock monitor.
• Call Manufacturer.
• Check version numbers.
• Call Manufacturer.
• Wait
• None
• Wait
• Check sample line and replace if
necessary.
• Check watertrap, replace or install,
if necessary.
• Check the connection and recon-
nect if necessary.
• If message persists, call Manufac-
turer.
• Wait
Scio MODULE D011100
Scio Module
78 68 123 E530U
78 68 115 E530U
59 53 539 E530U
Page 21