Status Messages

Message
Condition
Agent?
The module has not yet identified or
cannot identify agent because :
• Agent is unknown (ie., not HAL,
• Agent concentration is too low
• Vaporizer is leaking
• Traces of disinfectant are present
MultiGas Too Warm Fan port blocked
Hardware problem
i [parameter] > #
The inspired or expired concentra-
et [parameter] < #
tions of the parameter fall outside
the current upper or lower alarm
limits for that parameter
i [parameter]
The inspired concentrations of the
out of range (high)
parameter fall outside the moni-
i [parameter]
tor's measuring range
out of range (low)
et [parameter]
The expired concentrations of the
out of range (high)
parameter fall outside the moni-
et [parameter]
tor's measuring range
out of range (low)
MultiGas Data
Communication problem
Invalid
Mixed Agents
Scio Module agent identifier
detected the presence of more than
one anesthetic agent in the patient
circuit.
MultiGas Zero
Zero successful
Accepted
MultiGas Zero
Zero performed using the wrong
Failed
gas
Occlusion or leak present
HW problem
MultiGas Warming
Scio module is warming up and is
Up
operating at reduced accuracy
(? in parameter box)
Page 20
DES, ISO, SEV or ENF)
Scio MODULE D011100
Suggested Action
• Select agent manually (see
page 14).
• Check/replace vaporizer.
• Clear/unblock port.
• Call Manufacturer.
• Observe the patient and treat if
necessary.
• Adjust the alarm limits for that
parameter.
• Observe the patient and treat if
necessary.
• Check connections.
• Unplug and re-plug Scio module.
• Power-cycle monitor or un-dock
and re-dock monitor.
• Call Manufacturer.
• Unplug and re-plug Scio module.
• Power-cycle monitor or un-dock
and re-dock monitor.
• If message persists, call Manufac-
turer.
• If unintentional, check anesthesia
delivery system.
• After a change of agent, wait until
all traces of the first agent com-
pletely clear out of the breathing cir-
cuit.
• None
• Verify surrounding environmental
atmosphere is not contaminated.
• Check for leaks and occlusions.
• Call Manufacturer.
• Do not rely on Scio Module values.
SC 7/8/9XL VF2.x