1. Disconnect the sampling tubing.
2. Remove the trap from its connector by
holding it firmly on the ridged surfaces and
pulling it out from the Scio module.
3. Connect an empty syringe (size > 20 ml and
without a needle) to the port on the back of
the water trap.
4. Aspirate water trap contents into syringe.
5. Remove syringe and discard.
NOTE: Do not attempt to clean water trap.
6. Reseat water trap by pressing it against the Scio. A click will
indicate that it is seated properly.
7. Reconnect the sampling tubing. (Before monitoring a new
patient, exchange the tubing and T-connector, see page 17.)
Replacing the Water Trap
To replace the water trap:
1. Disconnect the sampling tubing.
2. Remove the trap from its connector by
holding it firmly on the ridged surfaces and
pulling it out from the Scio module
3. Insert the new trap by pressing it against the
Scio. A click will indicate that it is seated
properly.
4. Connect the sampling tubing. (Before monitoring a new
patient, exchange the tubing and T-connector, see page 17.)
Page 18
Scio MODULE D011100
SC 7/8/9XL VF2.x